Introduction

Calibrate the OMP40-2 spindle probe (P/N 34080-1): stylus runout, then length and diameter. To calibrate the OTS tool setting probe (P/N 34080-2), see VMCsi Probe Tool Option - Calibrate Tool Setting Probe.

Two paths, depending on what your shop has:

1-2-3 blocks only: Step 4, then Step 8 onward.

Precision Calibration Tool: Steps 5 to 7, then Step 8 onward.

Both paths end at the same calibration steps.

  1. Load the OMP40-2 into the spindle. Check runout at the ruby tip with a dial indicator while turning the spindle by hand. First pass: get runout within Β±0.0004 in.
    • Load the OMP40-2 into the spindle.

    • Check runout at the ruby tip with a dial indicator while turning the spindle by hand.

    • First pass: get runout within Β±0.0004 in.

    • Adjust the four lower flat set screws in opposing pairs.

    • Fine-tune with the two upper pointed set screws, also in opposition.

    • Final pass: reduce runout to Β±0.0001 in or better.

    • Re-adjust the lower screws as needed.

    • Finish with the upper screws. Do not calibrate until runout is within Β±0.0001 in.

  2. Pick the path for the equipment your shop has. Both paths end at Step 8.
    • Pick the path for the equipment your shop has. Both paths end at Step 8.

    • No calibration tool (1-2-3 blocks only):

    • Do Step 4, then skip to Step 8.

    • Precision Calibration Tool:

    • Do Steps 5 to 7, then Step 8.

  3. Stack two 1-2-3 blocks to make a 4 in fixed height on the table. Jog the spindle down until the blocks no longer slide underneath it.
    • Stack two 1-2-3 blocks to make a 4 in fixed height on the table.

    • Jog the spindle down until the blocks no longer slide underneath it.

    • Jog the spindle up slowly until the blocks just slide freely underneath.

    • The spindle should sit just above the blocks, not touching them.

    • Record the MZ1 value shown at this height.

    • You enter this value in Step 9. Skip to Step 8.

  4. Press [T, S, M]. Press [Work offset]. Select an available offset, for example G54.
    • Press [T, S, M].

    • Press [Work offset].

    • Select an available offset, for example G54.

    • Press [CYCLE START] to activate it.

    • Confirm the offset name (for example G54) shows on the screen.

    • Press [Select tool].

    • If the calibration tool is already in the tool list, select it, load it with [CYCLE START], and skip to Step 7.

    • If it is not, continue to Step 6 to add it.

  5. In the New tool - favorites window, press [Spec. tool 700-900]. Select 725 - Calibrating tool. Press [OK] to add it to the tool list.
    • In the New tool - favorites window, press [Spec. tool 700-900].

    • Select 725 - Calibrating tool.

    • Press [OK] to add it to the tool list.

    • Enter the tool length and diameter engraved on the calibration tool body.

    • Press [JOG] on the MCP.

    • Press [T, S, M], then press [Select tool].

    • Select the calibration tool and press [CYCLE START] to load it.

  6. Stack two 1-2-3 blocks on the table (4 in). Lower the calibration tool below the top of the blocks, beside them. Raise the tool slowly until the blocks just slide underneath without contact.
    • Stack two 1-2-3 blocks on the table (4 in).

    • Lower the calibration tool below the top of the blocks, beside them.

    • Raise the tool slowly until the blocks just slide underneath without contact.

    • Press [Act. values Machine] to turn off machine values. The display switches to work offset values.

    • Press [Z=0].

    • Confirm Z reads 0.

  7. Press [T, S, M]. Click the T field, then press [Select tool]. Select the 3D probe tool created during setup.
    • Press [T, S, M].

    • Click the T field, then press [Select tool].

    • Select the 3D probe tool created during setup.

    • Press [OK].

    • Press [CYCLE START].

    • Press [DOOR OPEN] to unlock the door.

    • Insert the OMP40-2 into the spindle.

    • Confirm the T field shows the 3D probe tool.

  8. Press [Meas. workp.]. Press [Calibrate probe]. Press [Length].
    • Press [Meas. workp.].

    • Press [Calibrate probe].

    • Press [Length].

    • No calibration tool: enter the MZ1 value from Step 4 in the Z0 field.

    • Calibration tool: enter 0 in the Z0 field.

    • Jog Z down until the probe is about 0.5 in above the 1-2-3 blocks.

    • Check the setup before starting the cycle.

  9. Press [CYCLE START]. The probe moves down, touches the top of the blocks, and measures its length.
    • Press [CYCLE START].

    • The probe moves down, touches the top of the blocks, and measures its length.

    • The "Calibrate: probe" window shows the machine values and trigger points.

    • Run the length calibration three times.

    • Confirm the machine values and trigger points repeat each run.

    • If they vary, lower the program feedrate and repeat.

  10. Remove the 1-2-3 blocks from the table. Place the ring gauge near the center of the table. Secure it with magnets or clamps so it cannot move.
    • Remove the 1-2-3 blocks from the table.

    • Place the ring gauge near the center of the table.

    • Secure it with magnets or clamps so it cannot move.

    • Jog so the stylus ball sits in the center of the ring gauge, below the rim.

    • In the "Calibrate: probe" window, press [Diameter].

    • Enter the ring gauge diameter in the ΓΈ field.

    • Press [CYCLE START].

    • The probe touches the inside of the ring gauge on all sides to measure diameter.

  11. When the cycle finishes, the window shows the machine values and trigger points.
    • When the cycle finishes, the window shows the machine values and trigger points.

    • Run the diameter calibration three times.

    • Confirm the machine values and trigger points repeat each run.

    • If they vary, lower the program feedrate and repeat.