Skip to main content
  1. If the Trunnion is not enabled, follow the procedure on the [TrakMT Options] screen.
    • If the Trunnion is not enabled, follow the procedure on the [TrakMT Options] screen.

    • [Setup] [TrakMT Options]

    • If the trunnion is being installed or its bolts are loose:

    • Position the trunnion in the correct mounting location.

    • Hand tighten the main mounting bolts so the trunnion is secured but still adjustable.

    • Lightly snug the adjustment block to hold position without fully locking it.

  2. Set the X-Axis to 12 o’clock and reset the A-Axis to 0.
    • Set the X-Axis to 12 o’clock and reset the A-Axis to 0.

    • Position the dial indicator ¼” from the back edge of the platter using the Y-Axis.

    • Using the Z-Axis in fine increments, preload the indicator to 0.004”, then set the dial pointer to 0.

    • Move the Y-Axis toward 6 o’clock until the indicator is ¼” from the near edge of the platter.

    • Check the dial indicator readings at the 12 o’clock and 6 o’clock positions:

    • If both readings are 0 and the A-axis display on the control panel reads 0.000°, skip to Step 5 (the A-axis is already referenced correctly).

    • If they differ, proceed to Step 3.

  3. Tilt the A-axis in fine increments until the dial indicator is midway between the start and end readings.
    • Tilt the A-axis in fine increments until the dial indicator is midway between the start and end readings.

    • For example, if the start reading is 0 and the end reading is 0.002”, adjust the A-axis until the indicator reads 0.001”.

    • Reset the dial indicator to 0 and move the Y-axis back to 12 o’clock on the rotary table.

    • Check if the dial indicator still reads 0 (or within 0.0001”).

    • If the readings at both 12 o’clock and 6 o’clock are 0, the A-Axis zero position is found.

    • To set this position as the new 0, proceed to Step 4 for A-Axis referencing.

  4. Step 4 is required only if you are adjusting and found 4th-Axis zero position from Step 3. Set Manufacture Password.
    • Step 4 is required only if you are adjusting and found 4th-Axis zero position from Step 3.

    • Set Manufacture Password.

    • Go to [Setup][Mach. data][Axis MD][Select Axis][AX4:MA1][Search] for 34210.

    • Turn on the [JOG], [REF. POINT], and [4th Axis] buttons on the MCP.

    • Change Parameter 34210 to "1"

    • Press the "[+]" button on the MCP. The value for Parameter 34210[0] should change from "1" to "2" automatically.

    • The changing of the value to "2" means that the referencing process is complete.

    • For Reference: VMCsi - Reset Machine Absolute Zero for Each Axis

    • Move the dial indicator to 3 o’clock on the platter, positioning it ¼” from the edge.

    • Preload the indicator to 0.004” and set the dial pointer to 0.

    • Sweep the X-axis from 3 o’clock to 9 o’clock across the platter.

    • Observe the indicator reading at 9 o’clock to determine if deviation exceeds 0.0008”.

    • If adjustment is required, loosen the cradle support bolts on the high side to lower the cradle into the correct position.

    • If that does not correct the issue, shim the trunnion on the low side using twice the measured difference. After shimming, restart the process from Step 2.

  5. Move A-axis to 0.000° (flat), mount a dial indicator to the spindle, and preload it on one edge of the platter slot.
    • Move A-axis to 0.000° (flat), mount a dial indicator to the spindle, and preload it on one edge of the platter slot.

    • Jog the X-axis across the slot (right to left) and rotate the C-axis until both ends read the same value. Record this C-axis position.

    • Tilt the A-axis to +90° or -90° (depending on Trunnion orientation) and repeat to obtain a second C-axis value.

    • Reverse the sign of both values, e.g., 0.082 → (–0.082), then calculate their average.

    • Go to [Setup][Mach. data][Axis MD][Select Axis][AX5:MC1][Search] for 34090[0].

    • If 34090[0] is 0.000, enter the averaged result directly.

    • If 34090[0] already contains a value, subtract the old value from the new averaged result and enter that correction.

    • Press [Reset PO] to apply the new reference offset.

  6. Tilt the A-axis to +90° or -90°, depending on the Trunnion’s orientation along the X-axis. Set up the dial indicator as shown in the image.
    • Tilt the A-axis to +90° or -90°, depending on the Trunnion’s orientation along the X-axis.

    • Set up the dial indicator as shown in the image.

    • In JOG mode, move the X and Z axes to position the indicator pointer vertically at the centerline of the rotary table, ¼” from the rightmost edge (3 o’clock position).

    • Slowly move the Y-axis to preload the indicator to 0.004”, then set the dial to zero.

    • Move the X-axis to the opposite (9 o’clock) side of the rotary table.

    • If the readings differ, loosen the left clamp and adjust the left side of the assembly along Y to correct.

    • Use adjustment screws 1 and 2 to fine-tune until deviation is ≤ 0.0001”.

    • Once aligned, tighten all clamps and torque both anchor blocks to 150 ft.lbs.

    • Position the dial gauge pointer as shown in the image.

    • Preload the indicator to 0.004” and set the dial pointer to 0.

    • Slowly rotate the platter 360 degrees, observing the + and - variations from zero.

    • Calculate the Total Indicator Reading (TIR) by finding the difference between the maximum and minimum values.

    • Ensure TIR does not exceed 0.002”.

    • If readings exceed the limit, inspect the platter mounting surface and flatness.

Finish Line

Ryan Levrini

Member since: 6/10/24

14 Guides authored

Team

Siemens Internal Member of Siemens Internal

54 Members

39 Guides authored

0 Comments

Add Comment



View Statistics:

Past 24 Hours: 0

Past 7 Days: 0

Past 30 Days: 2

All Time: 10